oil analyzer / oil / water / carbon
Organic X Monitor LC
Vnation JSC

-
Measured entity:
oil, oil, water, carbon, BTEX
-
Measured value:
trace, electrical conductivity, concentration, conductivity
-
Configuration:
for integration
-
Other characteristics:
with touchscreen, four-channel, automatic, online, IP65, IP54, with data logger, multi-channel, explosion-proof, dual-channel, ATEX, MCA, ppb, process, for hazardous environments, laboratory, ion
-
Technology:
UV
Continuous determination of inorganic and organic impurities in
? Condensates ? Boiler feed water ? Ultra pure water
One precondition for a continuous trouble free boiler operation is an almost perfect protective coating on both sides of the heat exchangers. The quality of these passivation layers is a function of the conductivity in the water - steam cycle, i. e. the concentration of ionic species.
Inorganic and organic ionic species are removed by ion exchange resins, and don't pose any thread to the protective layers. Neutral organic compounds are not removed and they are decomposed to organic and inorganic acids under conditions found in modern boilers. This yields to an increase in conductivity, higher corrosion and reduced maintenance intervals.
An analytical method that detects traces of organic impurities in feed water is therefore mandatory.